Millipore Sigma Vibrant Logo
Attention: We have moved. EMD Millipore products are no longer available for purchase on emdmillipore.com.Learn More

03-0210-00 SMC® Human pTau217 High Sensitivity Immunoassay Kit

03-0210-00
96 well assay  
Purchase on Sigma-Aldrich

Overview

Replacement Information
Description
Catalogue Number03-0210-00
Trade Name
  • SMC®
DescriptionSMC® Human pTau217 High Sensitivity Immunoassay Kit
OverviewThe SMC® Human pTau217 High Sensitivity Immunoassay Kit contains all reagents required to perform the assay. This assay takes advantage of a standard immunoassay workflow configured in a standard plate format. Assay protocols are similar to existing sandwich ELISA methods with two key differences: 1) Elution buffer disrupts the sandwich, separating the labeled detection antibody for quantification. 2) The SMCxPRO® immunoassay systems detect analytes using Single Molecule Counting (SMC®) technology.
Background InformationThe SMC® Human pTau217 High Sensitivity Immunoassay uses a quantitative fluorescent sandwich immunoassay technique to measure pTau217 in Human Serum, Plasma, and CSF samples. A capture antibody specific for human pTau217 has been pre-coated onto paramagnetic microparticles (beads). The user pipettes beads, standards, and samples into uncoated microplate wells. During incubation, the pTau217 present in the sample binds to the capture antibody on the coated beads. Unbound molecules are washed away during the subsequent wash steps. Fluor-labeled detection antibody is added to each well and incubated. This detection antibody recognizes and binds to pTau217 that has been captured onto the beads, thus completing the sandwich. Elution buffer is added to dissociate the protein sandwich, releasing the fluor-labeled antibodies. The eluted antibodies are transferred to a SMCxPRO® 384-well plate. The plate is loaded into the SMCxPRO® System where the labeled molecules are detected and counted. The number of fluor-labeled detection antibodies counted is directly proportional to the amount of pTau217 present in the sample. The amount of pTau217 in unknown samples is interpolated from a standard curve.
References
Product Information
Components
  • Assay Buffer;pTau217 Coated Beads;Standard Diluent;pTau217 Detection Antibody;pTau217 Standard;10X Wash Buffer;Buffer D;Elution Buffer B
Precision, %
  • The assay variations of SMC® Human pTau217 Immunoassay kits were studied using fifteen normal serum and plasma run in triplicate by 3 different operators on 3 different days.;Mean intra-assay variation was ≤10%.;Mean inter-assay variation was ≤20%.
Quality LevelMQ200
Applications
ApplicationThe SMC® Human pTau217 High Sensitivity Immunoassay Kit utilizes Single Molecule Counting technology, based on the trusted sandwich immunoassay principle, to digitally measure pTau217 in Human Serum, K2EDTA Plasma, and Cerebrospinal Fluid samples.
Key Applications
  • Single Molecule Counting (SMC®)
Application NotesTau protein is primarily a structural protein in neurons of the central nervous. Found as at least six different isoforms, these proteins are thought to provide structure, enable signaling, and regulate transport. In disease states, Tau proteins can become hyperphosphorylated and create tauopathies in the form of neurofibrillary or gliofibrillary tangles commonly associated with Alzheimer’s or Parkinson’s disease. Tau protein phosphorylated at threonine-217 is referred to as Phospho-Tau217 (pTau217). pTau217 is thought to be a promising marker for Alzheimer’s disease as it can be detected in serum or plasma and is strongly associated with physiological brain changes.
Biological Information
Cross ReactivityCross-reactivity to the following analyte was tested with the following result:
pTau181: Not detected
pTau231: Not detected
Accuracy
  • Five (5) samples each of serum, plasma, and CSF were spiked with three different concentrations of pTau217 standard. Spike Recovery Range: Serum 82 - 126%, Plasma 76 - 101%, CSF 77 - 94%; Spike Recovery Mean: Serum 102%, K2EDTA Plasma 94%, CSF 86%
Physicochemical Information
Sensitivity
  • Lower Limit of Quantification: 1.95 pg/mL
Dimensions
Materials Information
Toxicological Information
Safety Information according to GHS
Safety Information
Product Usage Statements
Usage Statement
  • Unless otherwise stated in our catalog or other company documentation accompanying the product(s), our products are intended for research use only and are not to be used for any other purpose, which includes but is not limited to, unauthorized commercial uses, in vitro diagnostic uses, ex vivo or in vivo therapeutic uses or any type of consumption or application to humans or animals.
Storage and Shipping Information
Packaging Information
Material Size96 well assay
Transport Information
Supplemental Information
Specifications
Global Trade Item Number
Catalog Number GTIN
03-0210-00 04065270902146