Quality Assurance and Process Control |
Analytical Quality Assurance becomes every day more important to secure the measure results of the samples. For waste and waste water analysis MilliporeSigma is the first supplier which offers diluted, ready-to-use concentrations of reference standards including the information about the expanded measurement uncertainty for photometric methods. Also standards for e.g. TOC analyzers, ICP-OES or Ion chromatography are offered ready-to-use to avoid dilutions.
Learn more about Certified Standard Solutions
Comprehensive quality assurance will transform your measurements with MilliporeSigma´s test systems into proven and verifiable analytical results. At MilliporeSigma, quality comes naturally to us. All our products are subject to diligent function and release testing. However, the user will still have to perform quality assurance in accordance with general laboratory standards. We recommend that you use our extensive range of process control options.
The quality assurance for MQuant®:
The concentration of these reference materials should correspond to the graduation which is close to the expected concentration of the sample. The most suitable standard solutions are described in the package insert for the relevant test kits. If the test yields the predetermined concentration of the standard solution, then this means that the entire analysis system is operating correctly.
To check the entire analysis system – test strips, instrument, pipettes and handling - the standard solutions are measured multiple times. Please note that the concentration of the standard solution should approximate the concentration of your sample. If the test yields the predetermined concentration of the standard solution within a tolerance of ± 10%, then this means that the entire analysis system is operating correctly. The most suitable standard solutions are described in the package inserts of the relevant test kits.
Find Standards
Foreign substances can interfere with the determination. To determine these interferences, there are two methods:
For the check of RQflex and RQflex plus we offer RQcheck Check Set
Instructions
Ordering Details
Spectroquant® NOVA 60A und Prove photometers have an instrument-supported Analytical Quality Assurance (AQA) and support your internal quality control.
Learn More
Pipettes should be tested at regular intervals in connection with Internal Quality Control measures (IQC). This is usually done by weighing the volume of a liquid on a calibrated precision balance. Such a balance is, however, not always readily available. With Spectroquant® PipeCheck you can check your pipettes and document the measurement results accordingly – even if you don’t have a balance.
Ordering Details
To make your analysis easier we provide not only a wide range of test-tools, but also multiple services for your support.