Chromolith® RP-18 Endcapped HPLC Columns
|
Ultra-High Performance on Any Instrument
Chromolith® 2 mm i.d. (internal diameter) columns for liquid chromatography offer an unparalleled combination of ultra-high performance and exceptionally low operating pressure. This leads to ultra-fast results, not only in the new UHPLC and UPLC instruments, but equally well in all standard HPLC systems with low dead volume.
The columns owe their rapid separation speed to their unique bimodal pore structure. The macropores have a diameter of 1.5 μm, resulting in a column efficiency that exceeds 100,000 plates/meter. Whereas the 13 nm (130 Å) mesopores are responsible for the large, uniform surface area on which adsorption takes place. Furthermore, Chromolith® RP-18 endcapped 2 mm i.d. columns have octadecylsilane surface modification, and are fully endcapped.
Features and Benefits
- Flexibility: Very fast, high-performance results with all low dead-volume LC instruments (UHPLC, UPLC, standard HPLC)
- Increased sensitivity and solvent efficiency: 81% solvents saved and 5.7 times higher sensitivity (compared to 4.6 mm i.d. columns)
- Flow rates from 0.2 - 1 mL/min: Compatible with LC-MS systems, and ESI or APCI interfaces
- Long column lifetime and resistance to blocking: Monolithic silica structure and absence of frits ensure method robustness and cost savings
Technical Data
Silica type |
High-purity |
Particle size |
Monolithic |
Macropore size |
1.5 μm |
Mesopore size |
13 nm (130 Å) |
Pore volume |
1 mL/g |
Total porosity |
> 80% |
Surface area |
300 m2/g |
Surface modification |
RP-18 endcapped |
Carbon content |
18% |
Chromolith® RP-18 Endcapped 2 mm i.d. Columns
Product
|
Internal Diameter
|
Ordering Number
|
Chromolith® Performance RP-18 endcapped |
100 x 2 mm |
152006 |
Chromolith® FastGradient RP-18 endcapped |
50 x 2 mm |
152007 |
Chromolith® Flash RP-18 endcapped |
25 x 2 mm |
152014 |
Application Examples
Chromolith® Performance RP-18 endcapped 100-4.6 mm
Column |
Chromolith® Performance RP-18 endcapped 100-4.6 mm |
Mobile Phase |
A: 100% Acetonitrile B: 100% Water + 0.1% TFA (v/v) C: 100% Methanol |
Isocratic |
Initial composition: A/B/C 30/60/10 (v/v/v) |
Flow rate |
2 mL/min |
Pressure |
45 bar (4.5 MPa, 65.3 psi) |
Detection |
Dionex Ultimate 3000 VWD-3400, 2.5 Hz, Response time 0.1 s, UV = 210 nm |
Vol. detector cell |
11 μL |
Temp. |
Ambient |
Injection volume |
1 μL |
Sample |
Bimatoprost Bimatoprost free acid |
|
Chromolith® Performance RP-18 endcapped 100-2 mm
Column |
Chromolith® Performance RP-18 endcapped 100-2 mm |
Mobile Phase |
A: 100% Acetonitrile B: 100% Water + 0.05% TFA (v/v) C: 100% Methanol |
Isocratic |
Initial composition: A/B/C 30/60/10 (v/v/v) |
Flow rate |
380 μL/min |
Pressure |
48 bar (4.8 MPa, 70 psi) |
Detection |
Dionex Ultimate 3000 VWD-3400, 2.5 Hz, Response time 0.1 s, UV = 210 nm |
Vol. detector cell |
1.4 μL |
Temp. |
Ambient |
Injection volume |
1 μL |
Sample |
Bimatoprost Bimatoprost free acid |
|
Separation of alkylphenones
Column |
Chromolith® Performance RP-18 endcapped 100-2 mm |
Mobile Phase |
A: Acetonitrile B: Water |
Gradient |
Time |
%ACN |
%Water |
0 |
15 |
85 |
3.5 |
90 |
10 |
5 |
90 |
10 |
5.1 |
15 |
85 |
6 |
15 |
85 |
Flow rate |
0.38 mL/min |
Pressure |
37-79 bar |
Detection |
254 nm |
Temperature |
Ambient |
Injection volume |
0.5 μL |
Sample |
1. Thiurea 2. Acetanilide 3. Acetophenone 4. Propiophenone 5. Benzophenone 6. Butyrophenone 7. Valerophenone 8. Hexanophenone 9. Heptanophenone in ACN/Water 60/40 |
|
Ultra-fast separation of antihistamines
Column |
Chromolith® FastGradient RP-18 endcapped 50-2 mm |
Mobile Phase |
A: 0.1% TFA in water B: 0.1% TFA in ACN |
Gradient |
5% to 90% B in 3.4 min |
Flow rate |
1.0 mL / min |
Pressure |
50 -120 bar |
Detection |
UV 230 nm |
Temp. |
Ambient |
Injection volume |
0.2 μL |
Sample |
1. Phenylephrine 2. Tripelenamine 3. Pyrilamine 4. Chloropheniramine 5. Brompheniramine 6. Chloropyramine 7. Diphenhydramine 8. Promethazine 9. Loratadine 10. Meclizine |
|
Carotenoids of salmon: Fast separation with Chromolith® Flash RP-18e 25-2 mm
Column |
Chromolith® Flash RP-18 endcapped 25-2 mm |
Mobile Phase |
A: Acetonitrile B: Water + 0.1% Formic acid |
Gradient |
Time |
%ACN |
%Water |
0 |
90 |
10 |
3 |
50 |
50 |
Flow rate |
1.14 mL/min |
Pressure |
22-53 bar |
Detection |
436 nm 11 μL flow cell |
Temperature |
Ambient |
Injection volume |
5 μL |
Sample |
1) Astaxanthin (cis + trans) 2) Canthaxanthin cis + trans) dissolved in acetonitrile/water + 0.1% formic acid 2/1 |
|
Bioflavonid separation: UHPLC with Chromolith® Performance RP-18e 100-2 mm
Column |
Chromolith® Performance RP-18 endcapped 100-2 mm |
Mobile Phase |
A: 0.1% TFA in H2O B: MeOH |
Gradient |
t [min] |
A [%] |
B [%] |
flow [mL/min] |
0.0 |
85 |
15 |
0.50 |
2.5 |
50 |
50 |
0.50 |
5.0 |
50 |
50 |
0.50 |
5.1 |
85 |
15 |
0.50 |
8.5 |
85 |
15 |
0.50 |
Detection |
220 nm UV |
Temperature |
Ambient |
Injection volume |
0.5 μL |
Sample |
1. Isoquerecetin 2. Troxerutin 3. Naringin 4. Morin 5. Querecetin 6. Trihydroxyethylluteolin |
|
Proteomics
Column |
Chromolith® Performance RP-18 endcapped 100-2 mm |
Mobile Phase |
A: 95% H2O/5% ACN/0.1% TFA (v/v/v) B: 5% H2O/95% ACN/0.085% TFA (v/v/v) |
Gradient |
from 5% B to 50% B in 20 min |
Flow rate |
0.3 mL/min |
Detection |
UV 214 nm |
Sample |
1 μL BSA digest (1mg/mL) |
|