pH Stability of Purospher® STAR RP-18 Endcapped


Robust Performance for the Most Demanding Applications

Purospher® STAR RP-18 endcapped is a robust and stable column suitable for a wide range of mobile phase conditions, which offers an extended lifetime. Furthermore, the columns provide the pH stability required for 99% of the most common analyses. Their outstanding pH stability has been demonstrated in numerous systematic studies, whether at pH 1.5 or 10.5. Their stability allows simple method development – even for the most demanding applications.

pH-Stability Tests

Purospher® STAR RP-18 endcapped columns were subjected to continuous operation in pH 1.5 or pH 10.5 mobile phases for more than 1300 hours. The stability of the column was evaluated by injecting two test mixtures. The virtually horizontal plots demonstrate the reproducibility of performance, capacity, and symmetry under acidic and basic conditions. The results of the chromatographic data prove the pH-stability of Purospher® STAR RP-18 endcapped under stringent chromatographic conditions.

Stability Test pH 1.5 und 10,5

Stability Test pH 1.5

Column Purospher® STAR RP-18 endcapped(5µm) LiChroCART® 150-4,6 HPLC column
Eluent Acetonitrile/Water (0.01% TFA; 50:50)
Flow rate 1.0 mL/min
Detection UV 254 nm
Temperature RT
Inj. volume 10 µL
Sample 1. Ketoprofen
2. Fenoprofen
3. Flurbiprofen
4. Ibuprofen
  Merck:/Freestyle/LE-Lab-Essentials/Chromatography/Diagrams/LE-Stability test at ph 1_5-201003.213.jpg

Stability Test pH 10.5

Column Purospher® STAR RP-18 endcapped
(5µm) LiChroCART® 150-4,6 HPLC column
Eluent Acetonitrile/Water
(0.1% NH3 (25%); 60:40)
Flow rate 1.0 mL/min
Detection UV 254 nm
Temperature RT
Inj. volume 10 µL
Sample 1. Thiourea
2. Benzene
3. Toluene
4. Propylbenzene
5. Butylbenzene
  Merck:/Freestyle/LE-Lab-Essentials/Chromatography/Diagrams/LE-Stability test at ph 10_5-201003.214.jpg

Stable Retention Times

The unique bonding technology of Purospher® STAR RP-18 endcapped provides an extremely stable stationary phase, ensuring excellent long-term stability. In more than 3000 sample injections of a test mixture at pH 10.5, retention times remained stable and no decline in performance was observed. This results in higher productivity at lower costs.

See more
Column
Purospher® STAR RP-18 endcapped (5µm) LiChroCART® 150-4,6 HPLC column
Eluent
Acetonitrile/Water (0.01% TFA; 50:50)
Flow rate
1.0 mL/min
Detection
UV 254 nm
Temperature RT
Inj. volume 10 µL
Sample 1. Ketoprofen
2. Fenoprofen
3. Flurbiprofen
4. Ibuprofen
  Merck:/Freestyle/LE-Lab-Essentials/Chromatography/Diagrams/LE-Stability Test at pH 10_5-201003.225.jpg

Determination of Betabion® at pH 10.1

p>Betabion® is a medicine for preventing vitamin B1 deficiency. During the HPLC method development for Betabion®, the pH of the mobile phase was found to have an effect on selectivity, and optimum selectivity was obtained at pH 10.1. This application demonstrates the excellent pH stability of Purospher® STAR RP-18 endcapped at high pH conditions, which is ideal for developing excellent separations and robust methods.

See more
Mobile phase gradient
Time 20 mmol/L
Tetraborate
buffer, pH 10.1 (%)
Acetonitrile (%) Flow (mL/min)
0 96 4 0.3
34 96 4 0.3
35 96 4 0.3
38 94 6 1,75
40 94 6 1,75
41 94 6 1,75
50 94 6 1,75
Column Purospher® STAR RP-18 endcapped (5µm) LiChroCART® 250-4,6 HPLC column
Flow rate 1.0 mL/min
Detection UV 240 nm
Temperature 22°C
Inj. volume 10 µL
Sample 1. Toxopyrimidine
2. Thiamin
3. 4-Methyl-5-thiazolethanol
  Merck:/Freestyle/LE-Lab-Essentials/Chromatography/Diagrams/LE-Determination of Betabion-300x345-11192013.jpg

 

 
Watch the Webcast: UHPLC2

UHPLC2 Two ways to speed up Chromatography

Learn More